Patient Monitor | PM-15D
- 15″ high resolution color TFT LCD display
- 13 Type Arrhythmic Analysis, Multi_Lead ECG Waveforms Display in Phase, Real time S_T segment analysis, pacemaker detection Drug calculation and titratiotable
- Efficient resistance to interference of defibrillator and electrosurgical cautery
- SPO2 can testing for 0.1% Weak
- RA-LL impedance Respiration
- Trend Coexist Display
- OxyCRG DyNamic View Display
- Bed to Bed view Display
- Networking capacity
- Options of printing, IBP, VGA and EtCO2
- UP to 4 hours working capacity of built_in rechargeable battery
- Large volume of tabular and graphic trends information storage and easy to recall
- Anti-ESU, anti- defibrillator
- Capture dynamic waveforms.
Specifications
ECG |
Lead Mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V) |
Lead selection: I, II, III, avR, avL, avF, V, |
|
Waveform: 2 ch |
|
Lead mode: 3 Leads ( R, L, F or RA, LA, LL) |
|
Lead selection: I, II, III, |
|
Waveform: 1 ch |
|
Gain: ′2.5mm/mV, ′5.0mm/mV, ′10mm/mV, ′20mm/mV, auto |
|
HR and Alarm |
|
Range Adult: 15 ~ 300 bpm |
|
Neo/Ped: 15 ~ 350 bpm |
|
Accuracy : ± 1% or ± 1bpm,which great |
|
Resolution: 1 bpm |
|
Sensitivity: > 200 (uV P-P) |
|
Differential Input Impedance: > 5 MΩ |
|
CMRR |
Monitor: > 105 dB |
Operation: > 105 dB |
|
Diagnosis: > 85 dB |
|
Electrode offset potential: ±300mV |
|
Leakage Current: < 10 uA |
|
Baseline Recovery: < 3 S After Defi. |
|
ECG Signal Range: ±8 m V ( Vp-p ) |
|
Bandwidth |
|
Surgery: 1 ~ 15 Hz |
|
Monitor: 0.5 ~ 35 Hz |
|
Diagnostic: 0.05 ~ 100 Hz |
|
Calibration Signal: 1 (mV p-p), Accuracy:±5% |
|
ST Segment Monitoring Range: Measure and Alarm -2.0 ~ +2.0 mV |
|
ARR Detecting |
|
Type: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, |
|
TACHY, BRADY, MISSED BEATS, PNP, PNC |
|
Alarm: Available |
|
Review: Available |
|
RESP |
Method: Impedance between R-F(RA-LL) |
Differential Input Impedance: >2.5 MΩ |
|
Measuring Impedance Range: 0.3~5.0Ω |
|
Base line Impedance Range: 0 – 2.5 KΩ |
|
Bandwidth: 0.3 ~ 2.5 Hz |
|
Resp. Rate |
Measuring and Alarm Range |
Adult: 0 ~ 120 rpm |
|
Neo/Ped: 0 ~ 150 rpm |
|
Resolution: 1 rpm |
|
Accuracy : ±2 rpm |
|
Apean Alarm: 10 ~ 40 S |
|
NIBP |
Method: Oscillometric |
Mode: Manual, Auto, STAT |
|
Measuring Interval in AUTO Mode: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min) |
|
Measuring Period in STAT Mode: 5 Min |
|
Pulse Rate Range: 40 ~ 240 bpm |
|
Alarm Type: SYS, DIA, MEAN |
|
Measuring and alarm range |
|
Adult Mode |
|
SYS: 40 ~ 270 mmHg |
|
DIA: 10 ~ 215 mmHg |
|
MEAN: 20 ~ 235 mmHg |
|
Pediatric Mode |
|
SYS: 40 ~ 200 mmHg |
|
DIA: 10 ~ 150 mmHg |
|
MEAN:20 ~ 165 mmHg |
|
Neonatal Mode |
|
SYS: 40 ~ 135 mmHg |
|
DIA:10 ~ 100 mmHg |
|
MEAN:20 ~ 110 mmHg |
|
Resolution Pressure:1mmHg |
|
Accuracy Pressure Maximum Mean error: ±5mmHg |
|
Maximum Standard deviation: ±8mmHg |
|
Overpressure Protection |
|
Adult Mode: 297±3 mmHg |
|
Pediatric Mode: 240±3 mmHg |
|
Neonatal Mode: 147±3 mmHg |
|
SpO2 |
Measuring Range: 0 ~ 100 % |
Alarm Range: 0 ~ 100 % |
|
Resolution: 1 % |
|
Accuracy : 70% ~ 100% ±2 % |
|
0% ~ 69% unspecified |
|
Actualization interval: about 1 Sec. |
|
Alarm Delay: 10 Sec. |
|
Pulse Rate |
|
Measuring and Alarm Range: 20~300bpm |
|
Resolution: 1bpm |
|
Accuracy : ±2bpm |
|
TEMPERATURE |
|
Channel: 2 |
|
Measuring and Alarm Range: 0 ~ 50 °C |
|
Resolution: 0.1°C |
|
Accuracy : ±0.1°C |
|
Average Time Constant:<10 Sec. |
|
IBP |
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2 |
ART: 0 ~ 300 mmHg |
|
PA: -6 ~ 120 mmHg |
|
CVP/RAP/LAP/ICP: -10 ~ 40 mmHg |
|
P1/P2: -10 ~ 300 mmHg |
|
Press Sensor |
|
Sensitivity: 5 uV/V/mmHg |
|
Impedance: 300-3000Ω |
|
Resolution: 1 mmHg |
|
Accuracy : ±2% or ±1mmHg, which great |
|
Actualization interval: about 1 Sec |
|
Standard: ECG, Respiration, NIBP, SpO2, Pulse Rate, Temperature |
|
Optional: Nellcor SpO2, EtCO2,IBP, Touch screen, Thermal Recorder,Wall mount, Trolley, Central station,VGA |